The Short Version

The QMSR builds on ISO 13485, but it adds FDA-specific requirements, and investigators can now ask for records they could not request before. A certificate shows that a registrar was satisfied. It does not show that your complaint files, supplier audits, and management reviews would hold up in front of an FDA investigator. This article covers where certified companies still have gaps, and how to find people who have worked under both systems.

The FDA’s Quality Management System Regulation took effect on February 2, 2026. It replaced the old Quality System Regulation in 21 CFR Part 820 and incorporates ISO 13485:2016 by reference. For a company that already holds an ISO 13485 certificate, that sounds like good news, and in part it is. A lot of the work is already done.

The trouble starts when “a lot of the work” becomes “all of it.” The QMSR is not simply the standard with a new name. It has its own definitions and a short list of FDA requirements that sit on top, and the FDA inspects differently than a registrar audits.

Quality professionals evaluating compliance requirements

A Certification Audit and an FDA Investigation Ask Different Questions

A registrar audits you against the standard, and the goal is a certificate. An FDA investigator inspects you against a federal regulation, and the goal is to decide whether your devices are made under adequate controls. Since February 2, 2026, FDA has done that using Compliance Program 7382.850, which replaced the Quality System Inspection Technique that investigators used for years.

The practical result is that the same procedure can pass one and draw questions in the other. A registrar may be satisfied that a process exists and is followed. An investigator is more likely to ask what the records show about how well it works, and what you did when it did not.

Four Places Certified Companies Still Have Gaps

1. Records that used to be off the table

Under the old regulation, FDA did not routinely inspect management review records, internal audit reports, or supplier audit reports. That exception is gone. FDA has stated that investigators now have the authority to review all three. Records created before February 2, 2026 can also be reviewed.

Companies with an ISO 13485 certificate already have these records, because registrars ask for them. The question is how they read to someone who is not a registrar. A management review that lists attendees but shows no decisions, or an internal audit with no follow-through on its findings, may have been acceptable for certification. It tells an investigator something different.

2. The FDA-specific requirements on top of the standard

The QMSR keeps a handful of provisions that ISO 13485 does not cover the same way. These include complaint records, unique device identification, and labeling and packaging controls. Complaint handling is a good example. ISO 13485 covers it in clause 8.2.2, and §820.35(a) of the QMSR then adds a requirement to keep records of the review, evaluation, and investigation of complaints involving the possible failure of a device, its labeling, or its packaging to meet specifications. For UDI, §820.35(c) requires the identifier to be recorded for each device or batch.

A quality system built only to the standard can miss these. A company can pass a certification audit and still have a complaint file that lacks something the FDA expects to see.

3. Definitions and terminology

The QMSR uses ISO 13485 and ISO 9000 definitions, with a few FDA exceptions. The familiar QSR records, the design history file, the device master record, and the device history record, are now described in ISO terms such as the design and development file and the medical device file. The content is still required. The names changed, and procedures and training materials that still use the old ones can confuse people during an inspection.

That is a smaller issue than the first two, but it is easy to overlook. It also tells you something about whether a team has worked through the change or only read about it.

4. Risk management that runs through the system

ISO 13485 expects risk management to be applied across the product lifecycle, not kept in a separate file. That means design changes, supplier selection, process validation, and post-market feedback should connect back to the same risk thinking. Companies that kept risk management as a stand-alone design activity often find the links missing when they look.

Records FDA can now request during an inspection Management review records, internal audit reports, and supplier audit reports were exempt from FDA inspection under the old regulation. Under the QMSR, FDA can review all three. UNDER THE QMSR, FDA CAN REQUEST Management review records Previously exempt Internal audit reports Previously exempt Supplier audit reports Previously exempt Records created before February 2, 2026 may also be reviewed.

An Example: A Certified Company Receives an Inspection Notice

Say a device maker has held its ISO 13485 certificate for years and has passed every surveillance audit. It receives notice of an FDA inspection. Three things are likely to happen that did not happen in its audits.

None of these is an unusual request. A company that has already looked at its system the way an investigator would can answer all three calmly. A company that has not may be reading its own records for the first time in front of the investigator.

Why a Certificate Alone Does Not Settle It

A registrar’s certificate is useful evidence that your system meets the standard. It does not replace an FDA inspection, and it does not address the parts of the regulation that go beyond the standard. It also says little about how your records will read to someone with different questions.

Common gaps worth checking include supplier files that do not document why a given level of control was chosen, management reviews that record attendance but not decisions and actions, and complaint procedures that do not address the FDA-specific record requirements. How much work each one takes depends on the company, and some may require a review of historical records. Either way, it is better to find them before an inspection than during one.

How to Tell Whether Someone Has Worked Under Both Systems

The people who help you close these gaps need to understand two systems, not one. Someone with twenty years under the old QSR may be new to ISO 13485. Someone who has only worked in ISO environments may never have been through an FDA investigation. Both are valuable, and both can leave a blind spot.

A short set of questions, asked the same way every time, tells you most of what you need. Listen for specific systems, records, and decisions. If an answer stays general, ask a follow-up.

Screening Questions
Which standards and regulations have you worked under? Ask for the systems and the roles, and whether the work covered both ISO 13485 and Part 820.
Tell me about a gap you found between the two. A good answer names a specific procedure or record and what changed.
How have your management reviews and internal audits read to an outside party? Ask what a reviewer could conclude from the records alone.
Describe a complaint file you had to bring up to standard. Listen for how the person handled evaluation, investigation, and reportability.
What did you change after an audit or inspection? Ask for the finding, the response, and the result.
Things to Weigh Carefully
Experience limited to one framework, with no sense of where the other differs.
Answers that describe the standard but never a record the person actually fixed.
No example of a document being questioned by a customer, registrar, or the FDA.
Reliance on the old QSR terms with no mention of the new ones.

Not every strong candidate has been through an FDA inspection. Registrar audits, customer audits, and supplier audits count too, as long as the person had to defend the work to someone outside the team.

How We Approach It at Bridgepoint GxP

We start with a gap assessment that compares your quality system with both the QMSR and ISO 13485, and then tells you what needs to change and in what order. Before a consultant joins a device client, we check that the person has worked with both frameworks, because experience in one does not guarantee experience in the other.

If you are screening on your own, the questions above are a good place to start. If you would rather have us do it, tell us about your situation and we will tell you how we would approach it.